Comparison of Central Blood Pressure Estimated by a Cuff-Based Device With Radial Tonometry
METHODS: Consecutive CuffCBP (SphygmoCor Xcel) and TonCBP (SphygmoCor 8.1) duplicate recordings were measured in 182 people with treated hypertension (aged 61±7 years, 48% male). Agreement between methods was assessed using standard calibration with brachial SBP and DBP (measured with the Xcel device), as well as with brachial mean arterial pressure (MAP; 40% form factor method) and DBP.
RESULTS: The mean difference ± SD for central SBP (cSBP), central DBP (cDBP), and central PP (cPP) between methods were -0.89±3.48mm Hg (intra-class correlation (ICC) 0.977; 95% confidence interval (CI) 0.973-0.982), -0.50±1.54mm Hg (ICC 0.992, 95% CI 0.987-0.993), and -0.42±3.57mm Hg (ICC 0.966, 95% CI 0.958-0.972), indicating good agreement. Wider limits of agreement were observed for central AP (cAP) and central AIx (cAIx) (-0.91±5.31mm Hg; ICC 0.802; 95% CI 0.756-0.839, -0.99±10.91%; ICC 0.749; 95% CI 0.691-0.796). Re-calibration with brachial MAP and DBP resulted in an overestimation of cSBP with CuffCBP compared with TonCBP (8.58±19.06mm Hg, ICC 0.164, 95% CI -0.029 to 0.321).
CONCLUSION:. cSBP, cDBP, and cPP derived from CuffCBP are substantially equivalent to TonCBP, although the level of agreement is dependent on calibration method. Further validity testing of CuffCBP by comparison with invasively measured central BP will be required.
History
Publication title
American Journal of HypertensionVolume
29Issue
10Pagination
1173-1178ISSN
0895-7061Department/School
Menzies Institute for Medical ResearchPublisher
Oxford University PressPlace of publication
United KingdomRights statement
Copyright 2016 American Journal of Hypertension, Ltd.Repository Status
- Restricted