eCite Digital Repository

Preliminary characterization of PRESAGE® for 3D dosimetry of 62 MeV proton beam

Citation

Gorjiara, T and Kacperek, A and Kuncic, Z and Baldock, C and Doran, S, Preliminary characterization of PRESAGE for 3D dosimetry of 62 MeV proton beam, Proceedings of the 7th International Conference on 3D Radiation Dosimetry (IC3DDose), 4-8 November 2012, Sydney, Australia, pp. 1-4. (2013) [Refereed Conference Paper]


Preview
PDF (Published manuscript)
352Kb
  

Copyright Statement

Licenced under Creative Commons Attribution 3.0 (CC BY 3.0) http://creativecommons.org/licenses/by/3.0/

DOI: doi:10.1088/1742-6596/444/1/012058

Abstract

PRESAGE® has previously shown potential for 3D dosimetry of heavy particles. A new formulation has specifically developed for dosimetry of protons/heavy ions. This work provides a preliminary characterization the new formulation of PRESAGE® by measuring optical absorbance and dose response after irradiating by a 62 MeV proton beam for a dose range of 0.5 – 20 Gy. Results show linear dose response and the evolution over time of the optical density of the 3D dosimeter.

Item Details

Item Type:Refereed Conference Paper
Keywords:three dimensional dosimetry, PRESAGE, radiological properties, water equivalency, Monte Carlo
Research Division:Physical Sciences
Research Group:Other Physical Sciences
Research Field:Medical Physics
Objective Division:Health
Objective Group:Other Health
Objective Field:Health not elsewhere classified
Author:Baldock, C (Professor Clive Baldock)
ID Code:97824
Year Published:2013
Web of Science® Times Cited:1
Deposited By:Office of the School of Engineering
Deposited On:2015-01-14
Last Modified:2017-11-01
Downloads:245 View Download Statistics