eCite Digital Repository
Photocured polymers in ion-selective electrode membranes .6. Photopolymerized lithium sensitive ion-selective electrodes for flow injection potentiometry
Citation
Farrell, JR and Iles, PJ and Dimitrakopoulos, T, Photocured polymers in ion-selective electrode membranes .6. Photopolymerized lithium sensitive ion-selective electrodes for flow injection potentiometry, Analytica Chimica Acta, 335, (1-2) pp. 111-116. ISSN 0003-2670 (1996) [Refereed Article]
DOI: doi:10.1016/S0003-2670(96)00355-8
Abstract
A photocured lithium ion-selective electrode, based on the ionophore N,N-dicyclohexyl-N',N'-diisobutyl-cis-cyclohexane-1,2- dicarboxamide (ETH 1810), was developed and evaluated. The robust nature of the photocured membrane made it ideally suitable for measurements in flow injection potentiometry within a linear range 0.1-10-3 M, detection limit of 5 x 10-4 M, and sample throughput of 150 h-1. The electrode was used successfully to determine lithium levels in pharmaceutical lithium carbonate tablets.
Item Details
Item Type: | Refereed Article |
---|---|
Research Division: | Chemical Sciences |
Research Group: | Analytical chemistry |
Research Field: | Analytical chemistry not elsewhere classified |
Objective Division: | Expanding Knowledge |
Objective Group: | Expanding knowledge |
Objective Field: | Expanding knowledge in the chemical sciences |
UTAS Author: | Dimitrakopoulos, T (Mr Telis Dimitrakopoulos) |
ID Code: | 7706 |
Year Published: | 1996 |
Web of Science® Times Cited: | 16 |
Deposited By: | Applied Science |
Deposited On: | 1996-08-01 |
Last Modified: | 2011-08-19 |
Downloads: | 0 |
Repository Staff Only: item control page