eCite Digital Repository

Testing Hygrometers Used in Cytogenetics Laboratories for Metaphase Preparation

Citation

Hartley, TF and Dun, K, Testing Hygrometers Used in Cytogenetics Laboratories for Metaphase Preparation, Current Protocols in Human Genetics, (Jul) pp. A.2F.1-A.2F.8. ISSN 1934-8258 (2011) [Refereed Article]


Preview
PDF
Restricted - Request a copy
228Kb
  

Copyright Statement

The definitive version is available at http://www.sciencedirect.com

DOI: doi:10.1002/0471142905.hga02fs70

Abstract

This protocol describes procedures for checking small laboratory hygrometers for accuracy at three relative humidity (rh) levels. The work arose out of the need to provide laboratory assessors with documentary evidence that the hygrometer used to monitor humidity in the vicinity of the laboratory where medical cytogenetics testing slides are prepared and dried in the ambient environment is reproducible and sufficiently accurate. The procedure is based upon the physicochemical principle that when water or certain saturated salt solutions are placed into a sealed environment, the humidity will equilibrate to well defined levels. We choose to check our hygrometers at three points: 95%, 75%, and 33% rh, using distilled water, saturated sodium chloride solution, and saturated magnesium chloride solution, respectively. Our results have demonstrated that the procedure is convenient and of sufficient accuracy to be fit for this annual hygrometer validation purpose. The procedure takes 24 hr per relative humidity point checked.

Item Details

Item Type:Refereed Article
Keywords:Hygrometer check - ISO 15189 - slide drying - environmental humidity - chromosome spreading - chromosome crossovers
Research Division:Medical and Health Sciences
Research Group:Clinical Sciences
Research Field:Pathology
Objective Division:Health
Objective Group:Health and Support Services
Objective Field:Diagnostic Methods
Author:Hartley, TF (Dr Thomas Hartley)
ID Code:75712
Year Published:2011
Deposited By:Health Sciences A
Deposited On:2012-02-10
Last Modified:2012-06-21
Downloads:0

Repository Staff Only: item control page