Using coupled monolithic rods for ultra-high peak capacity LC and LC-MS under normal LC operating pressures
Citation
Bones, J and Duffy, C and Macka, M and Paull, B, Using coupled monolithic rods for ultra-high peak capacity LC and LC-MS under normal LC operating pressures, The Analyst, 133, (2) pp. 180-183. ISSN 0003-2654 (2008) [Refereed Article]
DOI: doi:10.1039/b714807k
Abstract
Up to ten individual reversed-phase monolithic LC columns were efficiently coupled to equal a total column length of 1 m. The column generated up to 90 000 theoretical plates under optimum conditions with a peak capacity under gradient conditions of 244 peaks h(-1) at a total column backpressure of just 140 bar, directly coupled to ESI-MS.