Neutral analyte focusing by micelle collapse in partial-filling MEKC with UV and ESI-MS detection
Citation
Quirino, JP and Haddad, PR, Neutral analyte focusing by micelle collapse in partial-filling MEKC with UV and ESI-MS detection, Electrophoresis, 30, (10) pp. 1670-1674. ISSN 0173-0835 (2009) [Refereed Article]
![]() | PDF Restricted - Request a copy 121Kb |
Copyright Statement
The definitive published version is available online at: http://interscience.wiley.com
Official URL: http://interscience.wiley.com
DOI: doi:10.1002/elps.200800658
Abstract
Analyte focusing by micelle collapse is investigated as an on-line sample preconcentration technique in partial-filling MEKC with detection using UV and ESI-MS. Enhancement in peak height sensitivity of more than a 100 was obtained for model alkylphenylketones and dialkylphthalates using analyte focusing by micelle collapse-partial-filling MEKC with UV or ESI-MS detection.
Item Details
Item Type: | Refereed Article |
---|---|
Keywords: | Analyte focusing by micelle collapse • ESI-MS • Neutral analytes • On-line sample preconcentration • Partial-filling MEKC |
Research Division: | Chemical Sciences |
Research Group: | Analytical chemistry |
Research Field: | Separation science |
Objective Division: | Expanding Knowledge |
Objective Group: | Expanding knowledge |
Objective Field: | Expanding knowledge in the chemical sciences |
UTAS Author: | Quirino, JP (Associate Professor Lito Quirino) |
UTAS Author: | Haddad, PR (Professor Paul Haddad) |
ID Code: | 56721 |
Year Published: | 2009 |
Web of Science® Times Cited: | 32 |
Deposited By: | Austn Centre for Research in Separation Science |
Deposited On: | 2009-05-22 |
Last Modified: | 2010-04-08 |
Downloads: | 0 |
Repository Staff Only: item control page