Use of temperature programming to improve resolution of inorganic anions, haloacetic acids and oxyhalides in drinking water by suppressed ion chromatography
Citation
Barron, L and Nesterenko, PN and Paull, B, Use of temperature programming to improve resolution of inorganic anions, haloacetic acids and oxyhalides in drinking water by suppressed ion chromatography, Journal of Chromatography A, 1072, (2) pp. 207-215. ISSN 0021-9673 (2005) [Refereed Article]
DOI: doi:10.1016/j.chroma.2005.03.028
Abstract
Temperature programming was used to improve selectivity in the suppressed ion chromatographic (IC) separation of inorganic anions, haloacetic acids and oxyhalides in drinking water samples when using NaOH gradient elution. The programme exploited varying responses of these anions to changes in temperature. Heats of adsorption (Î"H, kJ/mol) for 17 anionic species were calculated from van't Hoff plots. For haloacetic acids, both the degree of substitution and log P (log of n-octanol-water partition coefficient) values correlated well with the magnitude of the temperature effect, with monochloro- and monobromoacetic acids showing the largest effect (Î"H = -10.4 to -10.7 kJ/mol), dichloro- and dibromoacetic acids showing a reduced effect (Î"H = -6.8 to -8.4 kJ/mol) and trichloro-, bromodichloro- and chlorodibromoacetic acids showing the least effect (Î"H = -4.7 to -2.4 kJ/mol). The effect of temperature on oxyhalides ranged from Î"H = 8.4 kJ/mol for perchlorate to Î"H = -9.1 kJ/mol for iodate. The effectiveness of two commercial column ovens was investigated for the application of temperature gradients during chromatographic runs, with the best system applied to improve the resolution of closely retained species at the start, middle and end of the separation obtained using a previously optimised hydroxide gradient, in a real drinking water sample matrix. Retention time reproducibility of the final method ranged from 0.62 to 3.18% RSD (n = 30) showing temperature programming is indeed a practically important parameter to manipulate resolution. © 2005 Elsevier B.V. All rights reserved.
Item Details
Item Type: | Refereed Article |
---|---|
Research Division: | Chemical Sciences |
Research Group: | Analytical chemistry |
Research Field: | Separation science |
Objective Division: | Expanding Knowledge |
Objective Group: | Expanding knowledge |
Objective Field: | Expanding knowledge in the chemical sciences |
UTAS Author: | Nesterenko, PN (Professor Pavel Nesterenko) |
UTAS Author: | Paull, B (Professor Brett Paull) |
ID Code: | 47587 |
Year Published: | 2005 |
Web of Science® Times Cited: | 31 |
Deposited By: | Chemistry |
Deposited On: | 2007-09-18 |
Last Modified: | 2011-08-18 |
Downloads: | 0 |
Repository Staff Only: item control page