Brachial and radial systolic blood pressure are not the same : evidence to support the Popeye phenomenon

Citation

Armstrong, MK and Schultz, MG and Picone, DS and Black, JA and Dwyer, N and Roberts-Thomson, P and Sharman, JE, Brachial and radial systolic blood pressure are not the same : evidence to support the Popeye phenomenon, Hypertension, 73, (5) pp. 1036-1041. ISSN 0194-911X (2019) [Refereed Article]

Copyright Statement

Copyright 2019 American Heart Association, Inc.

DOI: doi:10.1161/HYPERTENSIONAHA.119.12674

Abstract

Radial intra-arterial blood pressure (BP) is sometimes used as the reference standard for validation of brachial cuff BP devices. Moreover, there is an emerging wearables market seeking to measure BP at the wrist. However, radial systolic BP may differ when compared with brachial; yet some authors have labeled these differences as a fictional Popeye phenomenon. Indeed, differences between brachial and radial systolic BP have never been accurately quantified, and this was the aim of this study. Intra-arterial BP was measured consecutively at the brachial and radial artery in 180 participants undergoing coronary angiography (aged 6110 years; 69% men). On average, radial systolic BP was 5.5 mm Hg higher than brachial systolic BP. Only 43% of participants had radial systolic BP within 5 mm Hg of brachial. Additionally, 46%, 19%, and 13% of participants had radial systolic BP >5, between 5 and 10, and between 10 and 15 mm Hg higher than brachial respectively. A further 14% of participants had radial systolic BP >15 mm Hg higher than brachial, representing the so-called Popeye phenomenon. Finally, 11% of participants had radial systolic BP >5 mm Hg lower than brachial. In conclusion, radial systolic BP is not representative of brachial systolic BP, with most participants having a radial systolic BP >5 mm Hg higher than brachial and many with differences >15 mm Hg. Therefore, validation testing of BP devices utilizing intra-arterial BP as the reference standard should use intra-arterial BP measured at the same site as the brachial cuff or wearable device.

Item Details

Item Type:Refereed Article
Keywords:hemodynamics, hypertension, technology
Research Division:Medical and Health Sciences
Research Group:Cardiorespiratory Medicine and Haematology
Research Field:Cardiology (incl. Cardiovascular Diseases)
Objective Division:Health
Objective Group:Clinical Health (Organs, Diseases and Abnormal Conditions)
Objective Field:Cardiovascular System and Diseases
UTAS Author:Armstrong, MK (Mr Matthew Armstrong)
UTAS Author:Schultz, MG (Dr Martin Schultz)
UTAS Author:Picone, DS (Mr Dean Picone)
UTAS Author:Sharman, JE (Professor James Sharman)
ID Code:133817
Year Published:2019
Web of Science® Times Cited:3
Deposited By:Menzies Institute for Medical Research
Deposited On:2019-07-11
Last Modified:2019-08-05
Downloads:0

Repository Staff Only: item control page