Ultrastructural integrity of human capsulotomies created by a thermal device

Citation

Hooshmand, J and Abell, RG and Goemann, K and Davies, PEJ and Vote, BJ, Ultrastructural integrity of human capsulotomies created by a thermal device, Ophthalmology, 125, (3) pp. 340-344. ISSN 0161-6420 (2018) [Refereed Article]

Copyright Statement

Copyright 2017 by the American Academy of Ophthalmology

DOI: doi:10.1016/j.ophtha.2017.09.018

Abstract

Purpose: To investigate the ultrastructural features of anterior capsulotomy performed with a thermal device, the precision pulse capsulotomy (PPC).

Design: Prospective, multicenter case series.

Participants: Consecutive patients undergoing capsulotomy with the PPC device.

Methods: Prospective study of patients undergoing capsulotomy with the PPC by 2 surgeons, followed up by routine phacoemulsification cataract surgery, was undertaken. All capsulotomy specimens were collected for scanning electron microscopy (SEM). Observations were made regarding uniformity of the capsular edge and the presence of irregularities that may compromise integrity. Comparisons were made with manual continuous curvilinear capsulorrhexis.

Main Outcome Measures: Ultrastructural features of PPC and presence of irregularities.

Results: Frayed appearance of the anterior capsule edge was noted in postoperative visits under slit-lamp examination. Scanning electron microscopy sampling showed a generally uniform rolled capsular edge, but interspersed with areas of irregularity with frayed appearance at the capsule margin.

Conclusions: The PPC device is capable of creating reproducible, central, and precise circular capsulotomy. The ultrastructural features in ex vivo human capsulotomy specimens generally show eversion of the capsulotomy edge, but in some cases, this was accompanied by areas of irregular capsule margin with frayed edges, likely caused by dissipated thermal energy. The postoperative appearance and SEM features warrant further assessment of the PPC integrity and clinical correlation.

Item Details

Item Type:Refereed Article
Keywords:capsulotomies, PPC device
Research Division:Medical and Health Sciences
Research Group:Ophthalmology and Optometry
Research Field:Ophthalmology
Objective Division:Health
Objective Group:Clinical Health (Organs, Diseases and Abnormal Conditions)
Objective Field:Hearing, Vision, Speech and Their Disorders
UTAS Author:Goemann, K (Dr Karsten Goemann)
UTAS Author:Vote, BJ (Dr Brendan Vote)
ID Code:131020
Year Published:2018
Web of Science® Times Cited:8
Deposited By:Central Science Laboratory
Deposited On:2019-02-26
Last Modified:2019-04-15
Downloads:0

Repository Staff Only: item control page