University of Tasmania
Browse

File(s) not publicly available

A Photo-Cured Coated Wire Potassium Ion-Selective Electrode for Use in Flow Injection Potentiometry

journal contribution
posted on 2023-05-16, 10:56 authored by Alexander, PW, Dimitrakopoulos, T, Hibbert, DB
A potassium ion selective electrode (ISE) is reported based on the epoxyacrylate polymer prepared from o-nitrophenyl octyl ether as the plasticizer and 2,2′-bis [3,4-(15-crown-5-)-2-nitrophenylcarbamoxymethyl]tetradecane (BME-44) as the ionophore. Several different electrodes were studied by photo-curing different membranes onto a silver wire (0.8mm i.d.), and the response characteristics were evaluated. The membranes prepared exhibited strong adhesion to the metal substrate and showed improved mechanical strength compared to poly(vinyl chloride)-based membranes. The sensor exhibited Nernstian response over a log-linear range of between 0.01 mM and 100mM, and was selective against the common interfering ions. Fast response reaching 90% of steady-state value in 5 s makes this sensor applicable for flow injection potentiometry measurements. The potassium ISE allows for accurate determinations of potassium in various water samples by flow injection potentiometry.

History

Publication title

Electroanalysis

Volume

9

Issue

11

Pagination

813-817

ISSN

1040-0397

Department/School

University College

Publisher

VCH Verlagsgesellschaft mbH

Place of publication

Weinheim

Repository Status

  • Restricted

Socio-economic Objectives

Expanding knowledge in the chemical sciences

Usage metrics

    University Of Tasmania

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC